Test #4

Explore the possible improvement in analysis with line scan method.

With a video quality like this, line scan method does not gain significant improvement.

Video Source

CS-Alone.avi (generously provided by Dr. Javier Medina in the University of Pennsylvania.)

Download

EyelidTracker_v03.02.ipf

Results

Condition: Threshold = 60

Result

Line Scan vs. Frame Index (without thresholding) Line Scan vs. Frame Index (threshold = 60)
Line Scan vs. Frame Index Line Scan vs. Frame Index

Comparison between method #1 (thresholding and pixel counting: red) and method #2 (thresholding, line integration, and line counting: blue)
These two methods generated similar results.
Comparison

Video during Analysis

Summary

With a very good video quality, line scan method does not gain significant improvement.

One thing should be pointed out about the selection of ROI in the thresholding and pixel counting method (#1). Selecting the whole eye might not be the best way to determine ROI since the reflection of light source could introduce some error. Selecting a rectangle avoiding the reflecting spot as the ROI is better.